Application range and technical parameters of superimposed force standard machin

The microcomputer-controlled superimposed force standard machine is used for the measurement of standard dynamometers, load cells, and load sensors.
The main technical parameters of the microcomputer-controlled superimposed force standard machine:
Technical parameters of Dali value 0.1-level pressure-adding force standard machine:
2.1. Maximum test force: 5MN
2.2, Test force measurement range: 2%-100%F.S;
2.3. Force value error (advance and return):
2.4 Repeatability of force value:
2.5, initial load discrimination: at 10% FS, better than 0.02%
2.6. Load fluctuation:
2.7. Orientation error:
Within 2%-10%F.S, the test data is better than ±0.1%;
In the range of 10%-100%F.S, the test data is better than ±0.03%
In the range of 2%-10%F.S, the test data is better than 0.1%;
10%-100%F.S range; test data is better than 0.03%
In the range of 2%-10%F.S, the test data is better than 0.01%;
Test data in the range of 10%-100%F.S is better than 0.003%
In the range of 2%-10%F.S, the test data is better than 0.1%;
In the range of 10%-100%F.S, the test data is better than 0.06%